Table 1.

HPLC-ESI-MS conditions for analysis of C60 fullerene.

Chromatographic conditions
Column Agilent Eclipse XDB-C8
Column temperature 40 °C
Mobile phase methanol:toluene (45:55, v-v)
Flow rate 0.7 ml min
Run time 5 min
Injection volume 3 µl



MS conditions
Ionization source ESI
 desolvation line temperature 250 °C
 heat block temperature 400 °C
Target molecular ion 720 [M]+m/z
Time window 0–5 min
Dwell time 0.2 s
Interface voltage 4.5 kV
Nebulizing gas flow 3 l/min
Drying gas flow 15 l/min