Table 3.

Sensitivity and specificity for each haemoglobin phenotype identified using the reference standard method (Capillary Electrophoresis) in Lomé (Togo), May–June 2016. (N = 295)

Haemoglobin phenotype identified by Capillary Electrophoresis N Sensitivity (%) 95% CIa Specificity (%) 95% CIa
AA 86 100 [93·8–100] 100 [97·4–100]
AS 45 95·6 [84·8–99·5] 99·6 [97·8–99·9]
AC 39 94·9 [82·7–99·4] 99·2 [97·2–99·9]
SS 41 97·6 [87·1–99·9] 99·6 [97·2–99·9]
SC 44 97·7 [88·0–99·9] 99·6 [97·8–99·9]
CC 40 100 [87·1–100] 100 [97·8–100]

a 95% Confidence Intervals (CI) have been computed using the binomial distribution