Skip to main content
. Author manuscript; available in PMC: 2019 Jun 1.
Published in final edited form as: Microchem J. 2018 Mar 4;139:242–249. doi: 10.1016/j.microc.2018.03.010

Table 4.

Dry mass concentrations of Cd determined in SRM 3280 by ID-ICPMS along with residual solution concentrations of Mo and Sn measured in different of Scheme 3. Values are mean ± standard deviation of five separate preparations (n = 5). Values in parenthesis indicate per cent deviation (inaccuracy) from the certified value of Cd and those for Mo and Sn reflect per cent removal of Mo and Sn from SRM 3280 solution.

Step Analyzed medium Cd (ng g−1) Residual Mo (μg mL−1) Residual Sn (μg mL−1)

110Cd/113Cd 111Cd/113Cd 114Cd/113Cd

1 Supernatant 129 ± 10 136 ± 9 119 ± 12 1.27 ± 0.30 0.110 ± 0.012
(+61%) (+70%) (+49%) (76.0%) (86.7%)
2 Pellet 102 ± 8 112 ± 11 95.6 ± 7.1 0.32 ± 0.08 0.053 ± 0.006
(+28%) (+40%) (+20%) (93.9%) (93.6)
3 Pellet 78.8 ± 4.7 77.9 ± 5.2 76.5 ± 4.8 0.018 ± 0.008 0.019 ± 0.004
(−1.5%) (−2.6%) (−4.4%) (99.6%) (97.7%)