Outcome parameters

VivaSight Conventional intubation P
First-attempt success rate 96% 93% n.s.
Total number of attempts 1 attempt: 25 (96%)
2 attempts: 1 (4%)
1 attempt: 25 (93%)
2 attempts: 2 (7%)
n.s.
Average number of attempts 1.0 1.1 n.s.
Time to successful intubation (s) 34 (28–39) 35 (28–40) n.s.
Time to successful intubation with one attempt (s) 33 (28–39) 33 (28–38) n.s.
Vomiting/aspiration during intubation None None n.s.
Accidental oesophageal intubation None None n.s.
SpO2 < 80%
SpO2 (%)
19%
96 (82–99)
15%
99 (95–100)
n.s.
Systolic blood pressure < 70 mmHg
Systolic blood pressure (mmHg)
20%
93 ± 27
15%
95 ± 25
n.s.