TABLE 4.

Accuracy of the B1-Corrected, Variable Flip Angle, T1-Mapping Sequence for Each of the Phantom Vials

Phantom mean T1 (msec) 1 2 3 4 5 6 7 8
B1-corrected variable flip angle 351.3 351.3 1527 844.6 835.4 1661.4 1004 1526
Partial saturation recovery 350.6 346.8 1503 915.8 877.9 1691.9 999.8 1426
% Difference 0.2% 1.3% 1.6% 8.1% 5.0% 1.8% 0.4% 6.8%