Skip to main content
. 2018 Aug 15;4(8):FSO331. doi: 10.4155/fsoa-2018-0039

Table 1. . Intra-assay, interassay imprecision and inaccuracy of plasma paracetamol quantitation.

Nominal concentration (mg/l) Observed concentration (mg/l) Intra-assay Observed concentration (mg/l) Inter-assay

    Mean accuracy (%) SD %CV   Mean accuracy (%) SD %CV
0.125 0.122 86.6 2.0 1.8 0.122 86.3 2.5 2.3

0.119       0.117        

0.119       0.123        

0.123                

0.120                

0.25 0.251 98.9 2.3 0.9 0.251 99.1 2.3 0.9

0.254       0.250        

0.250       0.255        

0.253                

0.249                

5 5.282 110.0 55.4 1.0 5.282 110.1 40.0 0.7

5.226       5.224        

5.290       5.180        

5.215                

5.161                

30 32.920 109.4 334.5 1.0 32.920 109.3 462.9 1.4

32.465       31.978        

32.656       32.327        

32.086                

32.151             *  

%CV: Coefficient of variation.