Table 3.

Reproducibility of quantitative ultrasound technique

Study Year Equipment name/model Number of patients Intraobserver coefficient variant (%) Interobserver coefficient variant (%) Instrumental precision coefficient variant (%) Intersite variation coefficient variant (%)
Mercy et al. [2] 2007 Omnisense 7000P 84 1.26
McDevitt et al. [8] 2007 Omnisense 7000P 39 1.1 1.2
Zuccotti et al. [13] 2011 Omnisense 7000P 116 0.34
Gonnelli et al. [15] 2004 DBM Bone Profiler 140 1.0
McDevitt et al. [20] 2005 Omnisense 7000P 110 1.2 2.4
Rack et al. [23] 2012 Osteon KIV 172 0.62
Fewtrell et al. [25] 2008 99 1–2
Littner et al. [31] 2003 Omnisense 7000P 73 <1.2
Rubinacci et al. [32] 2003 DBM Sonic 1200 94 1.76 (standardised)
Littner et al. [34] 2005 Omnisense 7000P 22 <1.2
Liao et al. [38] 2010 Omnisense 7000P 267 1.23–1.84