TABLE 2.
Sensitivity and specificity of HS and 903 dried blooda
| Determinant | 903b |
HSc |
||
|---|---|---|---|---|
| Value (%) | 95% CI | Value (%) | 95% CI | |
| Sensitivity | 93.5 | 82.0–96.6 | 95.6 | 84.9–99.5 |
| Specificity | 100 | 85.2–100 | 100 | 85.2–100 |
| Positive predictive value | 100 | 100 | ||
| Negative predictive value | 88.5 | 72.0–95.8 | 92.00 | 74.8–97.8 |
| Accuracy | 95.7 | 87.8–99.1 | 97.1 | 89.8–99.6 |
HS and 903 spots were prepared from 30 reactive and 21 nonreactive clinical samples, plus 16 reactive and 2 nonreactive members from the SeraCare PRA204 low-titer panel.
For 903, 43/46 of the reactive plasma samples were reactive from 903 (3 were nonreactive), and 23/23 nonreactive from plasma were nonreactive from 903.
For HS, 44/46 of the reactive plasma samples were reactive from HS (2 were nonreactive), and 23/23 nonreactive from plasma were nonreactive from 903.