Table 4. Overall objective image quality for every scan protocol and kV setting.

Scan parameter p-value
High pitch (n = 81) Overall 70 kV 80 kV 90 kV
Coronary attenuation (HU) 367 ± 43 378 ± 48 358 ± 37 360 ± 38 0.141
Image noise 43 ± 6 43 ± 7 42 ± 6 45 ± 7 0.256
CNR 14 ± 4 14 ± 4 13 ± 3 14 ± 4 0.602
SNR 9 ± 2 10 ± 2 9 ± 1 9 ± 2 0.353
AS (n = 33)
Coronary attenuation (HU) 355 ± 37 354 ± 33 375 ± 52 336 ± 44 0.296
Image noise 40 ± 7 40 ± 6 38 ± 12 38 ± 6 0.676
CNR 16 ± 4 16 ± 4 17 ± 4 18 ± 2 0.635
SNR 10 ± 2 10 ± 2 11 ± 3 10 ± 2 0.261

Note–Values are presented as means ± standard deviation.

High pitch = prospective ‘high pitch’ scan protocol; AS = prospective ‘adaptive sequence’ scan protocol; SD = standard deviation; CNR = contrast-to-noise ratio; SNR = signal-to-noise ratio.