Published in final edited form as: J Pharm Biomed Anal. 2018 Mar 31;155:270–275. doi: 10.1016/j.jpba.2018.03.061

Table 2.

AZD3965 recovery, matrix effect, intra-day and inter-day accuracy and precision in mouse plasma. The measured concentrations represent the mean of triplicate measurements. The analyses were performed over 3 days.

Intra-day Inter-day Recovery Matrix
Effect




Standard
Curve
Nominal
concentration
(ng/mL)
Mean
(ng/mL)
SD Precision
(CV%)
Accuracy
(%)
Mean
(ng/mL)
SD Precision
(CV%)
Accuracy
(%)
% % SD
Low 10 10.3 0.06 0.56 103 10.1 0.225 2.23 101 101 103 4.18
150 153 4.93 3.22 102 148 4.36 2.94 98.9 101 -- --
400 415 8.33 2.01 104 402 17.7 4.39 101 102 108 1.93
High 1,000 1.04×103 15.3 1.47 104 1.01×103 30.1 2.98 101 103 101 1.84
6,000 6.00×103 124 2.07 100 5.96×103 144 2.42 99.4 101 -- --
10,000 9.76×103 262 2.68 97.6 9.57×103 161 1.68 95.7 105 97.8 1.41