Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2018 Aug 29;1097-1098:35–43. doi: 10.1016/j.jchromb.2018.08.027

Table 3.

Intra- and Inter-day accuracy and precision of the LC-MS/MS method for the quantification of CLBQ14 in various matrices. Values ≤ 20% for LLOQ and ≤ 15 % for other QCs are considered acceptable.

Biological Matrix QC Nominal Concentration (ng/mL) Intra-day (n=6) Inter-day (n=6)
Accuracy (RE, %) Precision (CV, %) Accuracy (RE, %) Precision (CV, %)
Plasma LLOQ 1 94.5 5.8 96.7 5.6
LQC 2.5 88.7 6.4 94.2 7.3
MQC 400 95.2 1.6 100.4 3.3
HQC 800 92.0 2.5 95.9 3.2
Urine LLOQ 1 96.1 7.6 100.9 7.2
LQC 2.5 99.0 9.2 96.4 4.7
MQC 400 99.6 2.3 100.6 2.6
HQC 800 97.9 2.6 99.8 2.5