Table 5:

Effect of dilution on the quantification of CLBQ14 using this LC-MS/MS method (n = 6). Values ≤ 15 % are considered acceptable.

Biological Matrix Dilution Factor Accuracy (RE, %) Precision (CV, %)
Plasma 5 99.7 3.5
10 94.6 8.6
20 95.6 10.7