TABLE I.

Example MR and Ultrasound Methods for Thermography and Thermometry.

Modality Method Temperature Dependence Tissue Dependence Sensitivity Spatial Res Temporal Res
MR Proton Resonance Frequency ω(θ) = γx0(1 − σ(θ)) Low ~0.01 ppm/°C
O(mm)
O(1s)
MR Spin-lattice Relaxation
T1~eEa(T1)kθ
High 1-2%/°C
O(mm)
O(1s)
MR Water Diffusion Coefficient
D~eEa(D)kθ
Medium ~2%/°C
O(mm)
O(1s)
US Mean Scatterer Resonance
f(θ)=c(θ)2ds(θ)
High ~0.05%/°C
O(mm)
O(ms)
US Echo Shift Equation 5 High ~0.05%/°C
O(mm)
O(ms)
US Backscatter Energy Equation 4 High 0.3 dB/°C
O(1cm)
O(1s)

Parameters: γ (gyromagnetic ratio), B0 (static magnetic field), σ (shielding constant), Ea (activation energy), k (Boltzmann constant); c (speed of sound), ds (mean scatterer spacing)