Published in final edited form as: J Magn Reson Imaging. 2018 Mar 31;48(5):1326–1335. doi: 10.1002/jmri.26025

Table 1.

Multiparametric MRI acquisition parameters given for prostate imaging at 3-Tesla with and without use of an endorectal coil.

Multiparametric MR Imaging Sequence Parameters at 3T

With endorectal coil Surface coil only

Parameter T2
Weighted
DWI* High b-
Value
DWI
DCE MR
Imaging
T2
Weighted
DWI** High b-
Value
DWI††
DCE MR
Imaging‡‡

Field of view (mm) 140 × 140 140 × 140 140 × 140 262 × 262 180 × 180 140 × 140 140 × 140 262 × 262
Acquisition Matrix 304 × 234 112 × 109 76 × 78 188 × 96 320 × 216 80 × 79 64 × 62 176 × 66
Repetition time (msec) 4434 4986 6987 3.7 3686 4766 7203 3.7
Echo time (msec) 120 54 52 2.3 120 45 47 2.3
Flip angle (degrees) 90 90 90 8.5 90 90 90 8.5
Section thickness (mm), no gaps 3 3 3 3 3 2.73 2.73 3
Image reconstruction matrix (pixels) 512 × 512 256 × 256 256 × 256 256 × 256 512 × 512 128 × 128 128 × 128 256 × 256
Reconstruction voxel imaging resolution (mm/pixel) 0.27 × 0.27 × 3.00 0.55 × 0.55 × 2.73 0.55 × 0.55 × 2.73 1.02 × 1.02 × 3.00 0.35 × 0.35 × 3.00 1.09 × 1.09 × 2.73 1.09 × 1.09 × 2.73 1.02 × 1.02 × 3.00
Time for acquisition (min:sec) 2:48 4:54 3:50 5:16 4:48 4:54 6:07 5:16
*

For ADC map calculation. Five evenly-spaced b values (0–750 sec/mm2) were used

b=2000 sec/mm2

DCE Images obtained, before, during, and after a single dose of gadopentetate dimeglumine 0.1mmol/kg at 3 mL/sec. Each sequence obtained at 5.6 s intervals

**

For ADC map calculation. Three evenly-spaced b values (0–600 sec/mm2) were used

††

b=1500 sec/mm2

‡‡

DCE Images obtained, before, during, and after a single dose of gadopentetate dimeglumine 0.1mmol/kg at 3 mL/sec. Each sequence obtained at 5.6 s intervals