Skip to main content
. 2018 Sep 28;11:337–343. doi: 10.2147/MDER.S180958

Table 2.

Days from index surgery to complaint by complaint type and device type. In some cases, surgery or complaint dates were not available.

Complaint type Device type
iFuse, days from index surgery to complaint
iFuse-3D, days from index surgery to complaint
N Mean SD Range N Mean SD Range
Allergy (metal) 2 151.5 204.4 7–296 0
Bone fracture 2 270.5 301.9 57–484 0
Cardiac incident 3 20.7 35.8 0–62 0
Embolism/aneurysm/DVT 3 24 16.1 11–42 0
Hematoma/seroma/bleeding 8 12 25.6 0–73 1 15
iFuse implant product problem 2 4.5 6.4 0–9 0
iFuse use problem 1 0 0
Infection 1 8 1 13
Instrument issue 31 126.5 362 0–1,529 0
Instrument use problem 0 0
Intraoperative issues 3 0.3 0.6 0–1 1 0
Off-label use 0 0
Others 1 965 0
Other medical procedures 5 609 177.8 413–819 0
Pain complaints (general) 151 520.3 376.8 3–1,651 3 41 52 2–100
Revision surgery 405 542.8 533.6 0–2,626 26 72.1 100.9 2–408
All 618 497.4 495.1 0–2,626 32 63.3 93.7 0–408

Note: Endashes indicate no value or statistic not calculable. In 4 cases, revision or index surgery dates were not known.