Table 6. Quantitative and qualitative evaluation of vessel continuity, vessel visibility and presence of artefacts in the SS-OCTA images.

Layer Parameters Pre-operative Post-operative P value
SCP Motion artefacts -54 8 .061
N of motion artefacts 4.77 ± 2.09 3.00 ± 2.94 .089
Image artefacts 23 85 .077
Distinguishability of FAZ -23 77 < .001*
Continuity of small vessels -38 33 .028*
Continuity of large vessels -46 62 .003*
N of bifurcations .923 ± .760 1.62 ± .290 .082
DCP Motion artefacts -31 23 .133
N of motion artefacts 3.15 ± 1.86 1.84 ± 1.91 .080
Image artefacts 23 85 .077
Distinguishability of FAZ -62 -8 .026*
Continuity of small vessels -77 23 .003*

FAZ = fovea avascular zone; SCP = superficial capillary plexus; DCP = deep capillary plexus; N = number.

The consensus grading contained an individual score for each feature of each image. Individual scores ranged from -1/0/+1, and were then summed up and normalized to a final score as presented in the Table. The maximal and minimal possible final scores ranged from -100 to +100. P-values were calculated by Pearson’s χ2 test. Number of motion artifacts and bifurcations were also analyzed and presented as means ± standard deviations in the Table. P-values were calculated by Student’s t-test.

*P < .05.

Continuity of large vessels and number of bifurcations were only graded in SCP images.