. 2018 Jun 13;99(8):1012–1026. doi: 10.1099/jgv.0.001083

Table 2. Time to positive (TTP) signal in the detection of different amounts of ZIKV synthetic RNA template 5 (from 5×106 to 5×102 copies per reaction) using the ZIKV RT-RPA assay.

For comparison, the time necessary to detect the target using the published Zika RT-PCR is also shown.

Template 5 (KU365780) copies/rxn RT-RPA TTP (min) RT-RPA result RT-PCR TTP (min) RT-PCR result
5×106 3.38 + (5/5) 40.12 + (3/3)
5×105 3.84 + (5/5) 45.07 + (3/3)
5×104 4.76 + (5/5) 49.18 + (3/3)
5×103 6.48 + (5/5) 53.38 + (3/3)
5×102 10.72 + (5/5) 57.25 + (3/3)
NTC Not detected − (5/5) Not detected − (3/3)