Table 1.
Comparative method (BD HLA‐B27 kit on BD FACSCalibur) | |||||
---|---|---|---|---|---|
Positive | Gray zone | Negative | BD FACSVia row total | ||
Test method (BD HLA‐B27 kit on BD FACSVia) | Positive | a = 72 | b = 8 | c = 0 | a + b + c = 80 |
Inconclusive | d = 1 | e = 11 | f = 3 | d + e + f = 15 | |
Negative | g = 0 | h = 0 | i = 499 | g + h + i = 499 | |
BD FACSCalibur column total | a + d + g = 73 | b + e + h = 19 | c + f + i = 502 | n = 594 |
Note: (a) Overall concordance was calculated as 100*{(a + e + i)/n} = 100*582/594 = 98%. (b) The five hundred and ninety four (594) samples were from 594 donors.