Skip to main content
. 2018 Apr 11;94(5):651–657. doi: 10.1002/cyto.b.21630

Table 3.

Summary Statistics for the Interlaboratory Reproducibility of BD FACSVia System

LMF statistics of HLA‐B27 FITC per study site
Donor type Study
Site
N Mean SD CV% Minimum Median Maximum
HLA‐B27
negative
BCW 30 137.13 1.87 1.36 134.00 137.00 141.00
MED 30 136.13 1.36 1.00 134.00 136.00 138.00
USC 30 134.40 1.73 1.29 130.00 134.00 138.00
HLA‐B27 positive BCW 30 173.00 1.62 0.94 170.00 173.00 176.00
MED 30 173.73 1.51 0.87 172.00 173.00 177.00
USC 30 172.23 2.36 1.37 167.00 173.00 178.00

Note: (a) BCW (BloodCenter of Wisconsin), USC (University of Southern California), MED (MedLab of BD Life Sciences). (b) CV% was calculated as SD/mean*100.