Table 4.

Precision Evaluation of the BD FACSVia System

Donor type Source of variation Total LMF mean SD with upper 95% CL
HLA‐B27 negative Within‐run repeatability 135.9 0.9 (1.0)
Total interlaboratory reproducibility 135.9 2.1 (2.4)
HLA‐B27 positive Within‐run repeatability 173.0 0.9 (1.0)
Total interlaboratory reproducibility 173.0 2.0 (2.3)

Note: (a) Combined data from three sites were used in the calculation of total LMF mean with SD (95% CL). (b) Total interlaboratory reproducibility included variables of within‐run, day and site/operator/instrument.