Skip to main content
. 2017 Nov 22;36(5):1508–1518. doi: 10.1002/jor.23771

Table 2.

Comparison of Different Press‐Fit Peg Types and the Optimum Level of Interference for Maximizing Pull‐Out Load

Peg Tapered Pull‐out Maximum ± S.D. (N) Radial Interference at Maximum (mm) Force Ratio (Out/In) at Maximum
Solid Yes 140 ± 4 0.8 0.17
No 77 ± 5 0.6 0.20
High Modulus Porous Yes 97 ± 6 0.8 0.15
No 65 ± 5 0.8 0.15
Low Modulus Porous Yes 102 ± 10 0.8 0.17
No 70 ± 2 0.8 0.17
Barbed Fixation (Baseline) No 433 ± 25 0.5 4.69