Published in final edited form as: Magn Reson Med. 2017 Jun 15;79(3):1595–1601. doi: 10.1002/mrm.26776

Table 1. Effect of the different correction methods on UCCA measures.

Top row: percent difference of measured UCCA at the extreme table positions (ZCImin and ZCImax). Bottom row: standard deviation of the group of measures performed with different table positions or at the different sites.

Correction
method:
None Scanner
2D
Scanner
3D
Phantom
fit
Vertebral
diameters
Jacobian
Healthy Control A
(single site)
15.2%
4.2mm2
6.8%
2.0mm2
2.1%
1.0mm2
2.9%
0.9mm2
1.9%
1.1mm2
0.7%
1.1mm2
Healthy Control B
(single site)
12.0%
3.6mm2
5.9%
2.0mm2
1.8%
1.2mm2
3.2%
1.3mm2
1.0%
1.9mm2
MS patient
(across 7 sites)
9.6%
2.5mm2
1.1%
1.6mm2
1.2%
1.7mm2