Skip to main content
. 2018 Apr 27;49(4):909–913. doi: 10.1016/j.bjm.2018.02.011

Table 2.

Comparison between LED-FM, GeneXpert sensitivity and specificity.

Total
Pleural fluid
BAL
CSF
Pus
Ascetic fluid
Synovail fluid
Urine
Pericardial Fluid
Bone Marrow
Tissues.
GX LED-FM GX LED-FM GX LED-FM GX LED-FM GX LED-FM GX LED-FM GX LED-FM GX LED-FM GX LED-FM GX LED-FM GX LED-FM
Sensitivity % 73 40 58 42 88 38 83 42 83 33 100 57 86 29 57 57 86 29 50 0 100 40
Specificity % 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
PPV 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
NPV 95 89 90 87 94 77 96 87 98 90 100 95 86 55 93 93 98 91 91 83 100 91

PPV, positive predictive value; NPV, negative predictive value; GX, GeneXpert; CSF, cerebrospinal fluid; LED-FM, light emitting diode fluorescent microscopy.