Table 2.

Summary of Main Characteristics of the Treatment Plans Concerning Dose Prescription and Arc Definition Settings.a

Breast Head and Neck Prostate
Dose prescription 2.67 Gy/fr; 15 fr SIB: 2.12Gy/fr; 33 fr SIB: 2.65 Gy/fr; 28 fr
PTV: 40.05 Gy PTV1: 54.4 Gy PTV1: 51.8 Gy
PTV2: 69.96 Gy PTV2: 65.5 Gy
PTV3: 74.2 Gy
2 arcs plans
Number or arcs 2 partial arcs 2 full arcs 2 full arcs
Coll. angle arc 1 (10/20)° (10/20)° 20°
Coll. angle arc 2 (350/340)° (350/340)° 340°
Gantry start angle ∼295° range: (285/307) Full rotation Full rotation
Gantry stop angle ∼173° range: (160/180)
Avoidance sector per arc Start: (345/20)°
Stop (90/110)°
3 arcs plans (Halcyon only)
Number or arcs 3 partial arcs 3 full arcs 3 full arcs
Coll. angle arc 1 (10/20)° (10/20)° 20°
Coll. angle arc 2 (350/340)° (350/340)° 340°
Coll. angle arc 3 90° 90° 90°
Gantry start angle ∼295°range: (285/307) Full rotation Full rotation
Gantry stop angle ∼173°range: (160/180)
Avoidance sector per arc Start: (345/20)°
Stop (90/110)°

Abbreviations: Coll., collimator; Fr, fractions; PTV, planning target volume; SIB, simultaneous integrated boost.

a The ranges expressed for the collimator rotation or in the start/stop angles for the breast and the head and neck plans represent the limits for the adjustment for individual patients.