Skip to main content
. Author manuscript; available in PMC: 2019 Oct 1.
Published in final edited form as: Top Magn Reson Imaging. 2018 Oct;27(5):319–333. doi: 10.1097/RMR.0000000000000177

Table.1.

Imaging parameters of a 2D-GRE-MRE for liver on a 1.5T magnet scanner

Parameters Typical values
Orientation Supine/Feet first
Coil Surface coil
Slice orientation Axial
Field of view (FOV) 30–42cm
Fractional FOV 0.7– 1.0 (patient specific)-optional
Frequency direction Right/Left
Matrix 256 ×64
Slice thickness 10mm
Number of slices 4
Breath hold Breath hold in expiration
TR/TE 50/20
Flip angle 25–30
Number of echoes 1
Echo train length 1
Bandwidth 31.25ms
Number of excitations (NEX) 1
Acceleration factor 2
Active driver frequency 60Hz
Active driver power 50% (up to 100% can be used)
Phase offsets 4
MEG frequency 60Hz
Number of MEG 1
MEG amplitude 1.76 Gauss/cm (variable and machine specific)
MEG direction R/L same as slice direction
Acquisition time 50–60s ( four breath holds)