Skip to main content
. 2018 Oct 11;13(10):e0205603. doi: 10.1371/journal.pone.0205603

Table 3. Categorical agreement of MBT-ASTRA (Bruker Daltonik GmbH) for AMX and CTX susceptibility testing of E. coli-positive blood cultures.

Sterile blood cultures inoculated with E. coli E. coli isolates retrieved from blood cultures
AMX test CTX test AMX test CTX test
Full agreement 93% 93% 95% 84%
Minor error Non applicable Non applicable Non applicable Non applicable
Major error 7% 7% 5% 16%
Very major error 0% 6% 2% 29%