Skip to main content
. 2018 Sep 14;45(10):4652–4659. doi: 10.1002/mp.13138

Table 2.

Dose rate and imaging dose metrics for XFCT and transmission CT configurations

XFCT Transmission CT
Incident X‐ray spectrum 125 kVp, 1.8‐mm Sn 62 kVp, 0.4‐mm Al
Dose rate at isocenter (cGy/min) 24.9 @ 24 mA 40.2 @ 1.0 mA
Acquisition time per projection (s) 10 0.05
Dose per projection (cGy) 1.87 (effective) 0.0335
Total number of projections 330 360
Total imaging dose (cGy) 617 (effective) 12.1