. 2018 Feb 28;17(4):582–587. doi: 10.1007/s12663-018-1098-z

Table 2.

Reconstructed area (%), raw roughness (Ra), and radiation dose (mSv) values for the bone specimen and each 3D model

Reconstructed area Roughness Radiation dose
Specimen 100.000A 2.288A
CB 0.2 89.042B 7.870BCD 2.450A
CB 0.25 89.797B 8.470D 2.380A
CB 0.3 58.930C 7.160BCD 1.230B
CT-16 0.625 99.880A 7.300BCD 9.900DE
CT-16 1.25 98.172A 6.330B 9.680CDE
CT-16 2.5 85.953B 6.970BCD 11.600E
CT-64 0.75 97.623A 7.140BCD 6.410CD
CT-64 1.0 98.875A 6.610BC 8.065CD
CT-64 1.5 96.032A 8.130CD 7.980CD

CB cone beam computed tomography, CT-16 fan beam computed tomography 16-channel, CT-64 fan beam computed tomography 64-channel. CT devices are followed by voxel size. Numbers followed by upper case letters indicate comparisons among the 3D models. Different letters indicate the statistical difference at p < 0.05; ANOVA followed by Tukey’s multiple comparison test