Skip to main content
. 2018 Jan 23;34(4):677–683. doi: 10.1007/s12288-018-0922-4

Table 3.

Characteristics of patient’s apheresis

Spectra Optia Fresenius Com.Tec p value
WBC × 103/µL 18.85 (1.7–82.3) 21.2 (2.1–63.0) p = 0.37
PLT × 103/µL 90.5 (16–229) 107 (33–300) p = 0.08
Procedure time (min) 227 (136–358) 174 (112–259) p = 0.001
Number of apheresis per case 2.3 (1–5) 2.3 (1–4) p = 0.63
ACD used (mL) 806 (336–1630) 580 (319–925) p = 0.001
Total blood volume (mL) 2.1 (1.3–4.8) 4.2 (2–5.2) p = 0.001
Product volume (mL) 177 (28–360) 220 (40–330) p = 0.001
Processed blood volume (L) 11,002 (3697–101,269) 7717 (4715–12,015) p = 0.07
Peripheral CD34+ cell count (µL) 46.68 (1.2–356) 49 (4.2–142) p = 0.77
CD34+ cell count in Product (µL) 1627 (32–8207) 1155 (0.2–6474) p = 0.018
CD34+ cells infused (106/kg) 3.69 (0.1–26.1) 3.3 (0.2–19.4) p = 0.58
Product WBC × 103/µL 280 (60–500) 212 (60–570) p = 0.02
Product Hct % 0.58% (0.1–2%) 1.17% (0.1–3%) p = 0.001
CD34+ collection efficiency 2 (CE2) % 89% (6–390%) 71% (3–200%) p = 0.07