Table 1. Mean and standard deviation (%ME, only for PA duration) of repeatability measurements for Fibion worn on the thigh (FT) and ActiGraph (95% Confidence Interval).
Sitting | Standing | Walking | |||||
---|---|---|---|---|---|---|---|
FT | ActiGraph | FT | ActiGraph | FT | ActiGraph | ||
AT | 1st test ME | 14.40 ± 0.30 (4.22) | 14.09 ± 1.26 (7.40) | 13.36 ± 1.03 (13.00) | 13.83 ± 1.12 (9.20) | 14.45 ± 0.45 (4.00) | 14.86 ± 0.78 (3.8) |
2nd test ME | 14.18 ± 0.33 (5.85) | 13.98 ± 0.95 (8.20) | 12.98 ± 0.90 (16.07) | 13.89 ± 1.41 (11.00) | 14.48 ± 0.39 (3.60) | 14.60 ± 0.88 (5.20) | |
CV% | 2.19 | 7.28 | 5.51 | 5.68 | 2.54 | 6.08 | |
p | 0.031 | 0.587 | 0.112 | 0.486 | 0.760 | 0.390 | |
ICC | 0.189 | 0.119 | 0.459 | 0.609 | 0.227 | −0.165 | |
EE | 1st test | 1.28 ± 0.201 | 1.30 ± 0.211 | 1.73 ± 0.23 | 1.74 ± 0.25 | 3.40 ± 0.61 | 3.52 ± 0.54 |
2nd test | 1.33 ± 0.18 | 1.38 ± 0.21 | 1.70 ± 0.25 | 1.76 ± 0.24 | 3.44 ± 0.73 | 3.47 ± 0.63 | |
CV% | 6.57 | 7.14 | 7.79 | 7.44 | 9.13 | 9.69 | |
p | 0.081 | 0.014 | 0.058 | 0.523 | 0.551 | 0.766 | |
ICC | 0.806 | 0.806 | 0.687 | 0.789 | 0.782 | 0.661 |
Note:
AT, refers to the duration of activity types; EE, refers to the energy expenditure; CV%, coefficient of variation; ME, percent measurement error; ICC, intra-class correlation coefficient; p-values refer to the Student’s t-test or the Wilcoxon signed-rank test.
The bold p-values indicating significant differences between the compared two groups.