Table 3. Pairwise differences and intraclass correlation coefficient (ICC) of FAZ measurements comparing the three different devices (R1, SCP, MD = mean difference, AD = absolute difference).

Device FAZ/mm2 Compared to MD AD p-value ICC
Optovue 0.309 ± 0.087 Canon 0.01 ± 0.02 0.02 ± 0.01 0.001 0.98
(0.953–0.991)
Heidelberg -0.02 ± 0.02 0.02 ± 0.02 0.000 0.964
(0.919–0.984)
Canon 0.295 ± 0.093 Heidelberg -0.03 ± 0.03 0.04 ± 0.02 0.002 0.958
(0.905–0.982)
Heidelberg 0.326 ± 0.089 - - - - -