Table 4:

Conversion factors for cardiac CT scans in 1-year-old phantom, based on dosimetry from 16-17 MOSFETs in the phantom, in three sub-experiments, using both ICRP Publication 103 and ICRP Publication 60 definitions of ED. The percentage difference for each scan is presented, comparing between a conversion factor derived using 50 MOSFETs per measurement (Table 2) vs. a conversion factor derived using 16-17 MOSFETs in three sets of measurements. All scans utilized a rotation time of 0.35 seconds, a small scan field of view (SFOV), scan length of 69.4 mm, and collimation of 64×0.625 mm. The scanner-reported DLP for all scans was based on the 32 cm CTDI phantom; n/a denotes not applicable.

Scan Description Tube
Potential
[kVp]
Pitch Conversion factor
ICRP 103
[mSv·mGy−1·cm−1]
% difference
from 50
MOSFETs
conversion factor
(Table 2)
Conversion factor
ICRP 60
[mSv·mGy−1·cm−1]
% difference
from 50
MOSFETs
conversion factor
(Table 2)
Axial 80 n/a 0.1001 1.79 0.0765 7.68
Axial 100 n/a 0.1012 3.85 0.0760 6.40
Axial with padding (100 ms) 80 n/a 0.1027 −6.22 0.0776 −1.20
Axial with padding (100 ms) 100 n/a 0.1011 −6.26 0.0755 −1.96
Helical 80 0.24 0.0991 −1.67 0.0787 −1.85
Helical 100 0.24 0.0931 −1.50 0.0745 0.27
Helical TCM * 80 0.24 0.0972 −0.89 0.0780 −0.25
Helical TCM * 100 0.24 0.0955 −0.58 0.0752 1.23
*

TCM denotes tube current modulation