Table 4:
Conversion factors for cardiac CT scans in 1-year-old phantom, based on dosimetry from 16-17 MOSFETs in the phantom, in three sub-experiments, using both ICRP Publication 103 and ICRP Publication 60 definitions of ED. The percentage difference for each scan is presented, comparing between a conversion factor derived using 50 MOSFETs per measurement (Table 2) vs. a conversion factor derived using 16-17 MOSFETs in three sets of measurements. All scans utilized a rotation time of 0.35 seconds, a small scan field of view (SFOV), scan length of 69.4 mm, and collimation of 64×0.625 mm. The scanner-reported DLP for all scans was based on the 32 cm CTDI phantom; n/a denotes not applicable.
Scan Description | Tube Potential [kVp] |
Pitch | Conversion factor ICRP 103 [mSv·mGy−1·cm−1] |
% difference from 50 MOSFETs conversion factor (Table 2) |
Conversion factor ICRP 60 [mSv·mGy−1·cm−1] |
% difference from 50 MOSFETs conversion factor (Table 2) |
---|---|---|---|---|---|---|
Axial | 80 | n/a | 0.1001 | 1.79 | 0.0765 | 7.68 |
Axial | 100 | n/a | 0.1012 | 3.85 | 0.0760 | 6.40 |
Axial with padding (100 ms) | 80 | n/a | 0.1027 | −6.22 | 0.0776 | −1.20 |
Axial with padding (100 ms) | 100 | n/a | 0.1011 | −6.26 | 0.0755 | −1.96 |
Helical | 80 | 0.24 | 0.0991 | −1.67 | 0.0787 | −1.85 |
Helical | 100 | 0.24 | 0.0931 | −1.50 | 0.0745 | 0.27 |
Helical TCM * | 80 | 0.24 | 0.0972 | −0.89 | 0.0780 | −0.25 |
Helical TCM * | 100 | 0.24 | 0.0955 | −0.58 | 0.0752 | 1.23 |
TCM denotes tube current modulation