Table 1.
Cable trap | 1 | 2 | 3 | 4 |
---|---|---|---|---|
64 MHz (S21 dB) | 48 | 50 | 47 | 45 |
128 MHz (S21 dB) | 42.42 | 47.18 | 39.15 | 26.22 |
The values shown in the table are the S21 measurement results in dB for each cable trap at 1.5T (64 MHz) and 128 MHz (3T) using a network analyzer and current probes. Each cable trap fulfilled the minimum requirement of 20 dB.