Skip to main content
. Author manuscript; available in PMC: 2019 Nov 1.
Published in final edited form as: J Thorac Cardiovasc Surg. 2018 Jun 2;156(5):1837–1848. doi: 10.1016/j.jtcvs.2018.05.042

Table 1.

Average hemodynamic parameters of the different valves. Values are reported ± standard deviation.

Valve Type Model Type ΔP (mmHg) EOA (cm2) RF (%) Pinwheeling Index Peak RSS (Pa)
Evolut Patient-Specific 13.9±0.19 1.5±0.004 32.10±0.70 0.201±0.05 161.27±3.45
Rigid 10.5±0.15 1.8±0.036 15.74±0.73 0.122±0.045 337.22±7.05
SAPIEN 3 Patient-Specific 5.0±0.09 1.7±0.011 23.75±0.75 0.231±0.057 122.84±1.76
Rigid 7.76±0.083 2.1±0.025 10.92±0.11 0.366±0.067 157.91±1.80