Skip to main content
. 2018 Mar 12;57(17):4752–4755. doi: 10.1002/anie.201800954

Table 1.

Logarithmic selectivity coefficients of gold nanoporous membrane based Cu2+‐selective electrodes.

Ion [J]
logKCu,Jpot
Cu2+‐selective nanopore peptide‐free nanopore conventional ionophore‐based ISEs
Na+ −5.56 −0.71 −3.7,[b] −4.61[d]
K+ −5.71 −0.30 −2.4,[a] −3.7,[b] −4.74[d]
Li+ −5.70 −0.84
Cs+ −5.76 0.16
H+ −2.07 −0.08 −0.44[d]
Ca2+ −3.63 −0.27 −4.3,[b] −3.3,[c] −4.64[d]
Mg2+ −3.76 0.11 −4.0,[b] −3.3,[c] −6.26[d]
Zn2+ −4.26 0.04 −2.3,[a] −3.9,[b] −3.0,[c] −2.67[d]
Cd2+ −3.19 0.07 −4.4,[a] −4.4,[b] −2.7,[c] −3.34[d]
Ni2+ −2.97 −0.10 −3.3,[a] −3.9,[b] −3.0[c]
Pb2+ −3.08 0.02 −0.8,[a] −1.8,[b] −2.7,[c] −1.82[d]
Cr3+ −4.37 −0.50 −3.0[c]

[a] o‐xylylene bis(dithiocarbamate,15 [b] tetraethylthiuram disulfide,16 [c] salicylaniline,17 [d] N,N,N′,N′‐tetracyclohexyl‐3‐thiaglutaric diamide.18