Skip to main content
. 2017 Jan 30;2(1):18–25. doi: 10.1080/20961790.2016.1270812

Table 2.

Inter- and intra-day imprecision and accuracy for quality controls at 10 and 100 ng/L.

  Imprecision (%)
Accuracy (%)
  Inter-day (n = 5)
Intra-day (n = 4)
Inter-day (n = 5)
Intra-day (n = 4)
Compound 10 (ng/L) 100 (ng/L) 10 (ng/L) 100 (ng/L) 10 (ng/L) 100 (ng/L) 10 (ng/L) 100 (ng/L)
Morphine 11.4 6.2 5.6 3.4 93.3 94.7 116.0 110.9
Oxymorphone 10.8 3.9 4.9 3.3 106.5 106.6 126.5 109.1
Hydromorphone 14.1 3.3 12.9 4.4 98.3 103.3 119.5 121.1
Oxycodone 8.9 7.6 10.9 5.0 98.3 101.1 110.0 102.0
Hydrocodone 10.0 8.8 6.1 4.8 107.5 103.7 131.0 127.2
THC 9.7 9.0 21.2 7.9 102.8 98.8 110.3 106.6
THCCOOH 8.0 6.1 5.0 7.3 97.3 105.6 119.3 103.4
THCCOOH-glucuronide 6.0 10.2 4.1 5.5 95.0 98.9 83.0 102.3