Skip to main content
. 2018 Oct 24;8:15700. doi: 10.1038/s41598-018-33628-z

Table 4.

Parameters and information for each specimen.

Specimen I II II IV Phantom
Tumor characteristics as verified by histopathology
Histological diagnosis multicentric lobular invasive carcinoma, G2, and lobular carcinoma in situ recurrent, invasive carcinoma of no specific type (NST, formerly invasive ductal), G3 invasive carcinoma of no specific type (NST, formerly invasive ductal), G2, and adjacent intraductal carcinoma (DCIS) bifocal invasive carcinoma of no specific type (NST, formerly invasive ductal), G1, and intraductal carcinoma (DCIS)
Max. tumor diameter 51 mm 47 mm 5 mm 25 mm
Sample position for ex-vivo mammography
Orientation AP AP AP CC
Compressed thickness [cm] 4.0 6.0 4.5 6.0 4.5
Clinical acquisition parameters
X-ray tube settings 27 kVp (W/Rh) 35 kVp (W/Ag) 30 kVp (W/Rh) 39 kVp (W/Ag) 28 kVp (W/Rh)
72 mAs 92 mAs 100 mAs 23 mAs 200 mAs
MGD civAC-Mx [mGy] 1.3 3.8 2.9 1.3
MGD cevAC-Mx [mGy] 0.9 2.2 1.4 1.1 2.0
Acquisition parameters MuCLS
Energy 25 keV 25 keV 25 keV 25 keV 25 keV
MGD mAC-Mx [mGy] 0.3 0.4 1.0–2.0
Exposure times [s] 3 3 10–20
MGD mgb-Mx [mGy] 0.8 0.9 0.9 0.7–1.8
Exposure times [s] 7 14 11 7–18
Stitching 4 × 4 5 × 5 5 × 5 4 × 4 2 × 2