Validation of linearity tested with Mandel test according to the method of Kromidas (8)

Analyte Concn range (mg/liter) Test value (Fcalc = Ds2/SyQ2)a Ftab value Proof of linearity
F901318 0.05–12.00 9.977 21.1977 Yes: y = 0.34986x + 0.038507; R2 = 0.99955
a

Ds2, variance difference; SyQ2, variance of quadratic regression fit.