Skip to main content
. 2018 Jul 10;89(5):490–495. doi: 10.1080/17453674.2018.1490985

Table 2.

Precision of radiostereometric analysis based on double examinations

  Translation (mm)
Rotation (°)
  x-axis y-axis z-axis x-axis y-axis z-axis
Clinical precision 0.13 0.22 0.36 0.53 1.95 0.19

Clinical precision =2.05 x standard deviation for the mean difference between the double examinations.