Table 2.
Final checks of the device after lead exchange
New RA lead | Tined passive lead (7Fr) |
Threshold PSA 0.50 @ 0.50 | |
Impedance PSA 536 | |
RV lead (unchanged) | Active fix 52 cm bipolar |
Threshold PSA 0.70 @ 0.50 | |
Impedance PSA 641 |
RA, right atrial.
Final checks of the device after lead exchange
New RA lead | Tined passive lead (7Fr) |
Threshold PSA 0.50 @ 0.50 | |
Impedance PSA 536 | |
RV lead (unchanged) | Active fix 52 cm bipolar |
Threshold PSA 0.70 @ 0.50 | |
Impedance PSA 641 |
RA, right atrial.