. 2018 Sep 18;45(10):4471–4482. doi: 10.1002/mp.13133

Table 1.

Summary of method performance averaged over the 19 subjects of the full projection dataset as quantified by the image quality measures of TIS at the diaphragm and bronchus intermedius and noise of three ROIs (air, aorta, and soft tissue). Percentage metrics presented represent the relative percentage of TIS recovered across the three evaluated interfaces relative to the initial 4D‐CBCT reconstruction

Reference frame and method
Average of 4D‐CBCT Fixed frame (ff); registration‐only Fixed frame (FF); MoCo recon. Mean frame (MF); registration‐only Mean frame (MF); MoCo recon.
Mean (%) Std. Dev. (%) Mean (%) Std. Dev. (%) Mean (%) Std. Dev. (%) Mean (%) Std. Dev. (%) Mean (%) Std. Dev. (%)
TIS recovery
Diaphragm 95.7 (39.0) 98.5 (20.0) 98.4 (19.9) 122 (43.8) 121 (43.4)
Bronchus intermedius 82.5 (20.7) 95.4 (9.24) 94.9 (10.3) 94.3 (20.4) 93.8 (21.8)
ROI noise reduction
Ambient air 64.4 (9.95) 64.1 (8.86) 7.6 (11.2) 62.7 (9.98) 67.6 (11.4)
Aorta 59.5 (6.82) 57.0 (5.86) 57.3 5.44) 59.0 (5.55) 58.2 (5.55)
Soft tissue 52.3 (11.7) 42.7 (8.41) 43.5 (9.42) 40.5 (8.64) 41.7 (8.95)