Skip to main content
. Author manuscript; available in PMC: 2019 Oct 26.
Published in final edited form as: Biomed Chromatogr. 2018 Apr 26:e4270. doi: 10.1002/bmc.4270

Table 1.

Summary of calibration curves

TFV FTC EVG RPV
Calibration curve range 10 to 2000 ng/mL
Inter-assay Accuracy (% RE) (n=3) −8.9 to 9.7 −8.1 to 7.6 −6.1 to 7.7 −3.3 to 5.6
Inter-assay Precision (% CV) (n=3) 0.5 to 8.1 0.07 to 8.2 0.19 to 4.2 1.11 to 4.8
Correlation coefficient (r) > 0.9961 > 0.9978 > 0.9988 > 0.9985