. 2018 Oct 25;56(11):e00992-18. doi: 10.1128/JCM.00992-18

TABLE 1.

Comparison of the ViraChip MIB assays and ViraStripe LIA for detection of anti-B. burgdorferi IgM and IgG antibodies in serum

ViraChip assay and resulta ViraStripe assay and no. to tests with indicated resultb
% agreement (95% CI)c
Kappa value (95% CI)
Positive Negative Uninterpretable Positive Negative Overall
IgM MIB IgM LIA (n = 411)
    Positive 160 1 1
    Negative 49d 191 4 75.5 (69.2–80.8) 99.0 (96.1–99.9) 85.6 (81.9–88.7) 0.71 (0.65–0.78)
    Invalid 3 1 1
IgG MIB IgG LIA (n = 411)
    Positive 122 25 0
    Negative 9 253 0 93.1 (87.3–96.5) 90.4 (86.3–93.3) 91.2 (88.1–93.6) 0.81 (0.75–0.87)
    Invalid 0 2 0
a

MIB, microarray immunoblot; LIA, line immunoassay.

b

The ViraChip IgM MIB assay detected antibodies to only p41 in 37/49 samples and to only p23 in only 1 sample. The ViraChip IgM MIB assay did not detect antibodies to p39 in any of the 49 discordant samples.

c

CI, confidence interval.

d

Eleven samples had 0 bands, and 38 samples had only 1 band detected by the ViraChip IgM MIB assay.