Skip to main content
. 2018 Jun 26;46(11):1745–1755. doi: 10.1007/s10439-018-2073-z

Figure 8.

Figure 8

Diameter and insertion depth errors. Left, diameter error (OCT–µCT and IVUS–µCT) of holes in OCD and CD. Right, insertion depth error (OCT–µCT and IVUS–µCT) of holes in OCD and CD. Drill bits had a diameter of 0.1, 0.5, 1, 2 and 3 mm. The aimed for drill depth was 4 mm for OCD and 1 mm for CD. “*” indicates a significant difference between populations.