Combined standard uncertainty for voltage measurement by the ADC component.

Source Value (V) Type Probability Distribution Standard Uncertainty (V)
Quantization error a 3 × 10−4 B Rectangular 1.7 × 10−4
Signal to distortion ratio b 2 × 10−4 B Rectangular 1.2 × 10−4
Repeatability c 1.5 × 10−4 A Normal 1.5 × 10−4
Combined standard uncertainty, uCount 2.5 × 10−4

a ±0.5 LSB (least significant bit) Analog Devices AD7490 12-bit ADC resolution 0.6 mV BL−1; b 74 dB; c experimental standard deviation of the mean of 30 independent measurements of five constant voltages.