Table 5.

Uncertainty budgets for the calibration stage.

Uncertainty Source Type 5 °C [°C] 25 °C [°C] 45 °C [°C]
Reference thermometer calibration B 1 × 10−3 2 × 10−3 5 × 10−3
Self-heating of thermistor B 7.3 × 10−8 2.9 × 10−8 1.2 × 10−8
Calibration model interpolation error a A 8 × 10−3 8 × 10−3 8 × 10−3
Continuous temperature sweep error A 2.3 × 10−3 2.3 × 10−3 2.3 × 10−3
Measurement noise of reference thermometer A 7.3 × 10−4 7.9 × 10−4 8.1 × 10−4
ADC count measurement A,B 8 × 10−3 8 × 10−3 8 × 10−3
Total combined standard uncertainty 1.2 × 10−2 1.2 × 10−2 1.3 × 10−2

a See estd in Table 8.