Table 5.
Uncertainty Source | Type | 5 °C [°C] | 25 °C [°C] | 45 °C [°C] |
---|---|---|---|---|
Reference thermometer calibration | B | 1 × 10−3 | 2 × 10−3 | 5 × 10−3 |
Self-heating of thermistor | B | 7.3 × 10−8 | 2.9 × 10−8 | 1.2 × 10−8 |
Calibration model interpolation error a | A | 8 × 10−3 | 8 × 10−3 | 8 × 10−3 |
Continuous temperature sweep error | A | 2.3 × 10−3 | 2.3 × 10−3 | 2.3 × 10−3 |
Measurement noise of reference thermometer | A | 7.3 × 10−4 | 7.9 × 10−4 | 8.1 × 10−4 |
ADC count measurement | A,B | 8 × 10−3 | 8 × 10−3 | 8 × 10−3 |
Total combined standard uncertainty | 1.2 × 10−2 | 1.2 × 10−2 | 1.3 × 10−2 |
a See estd in Table 8.