Parameter Value
Polarity Positive
Full MS
Microscans 1
Orbitrap resolution 60,000
AGC Automatic Gain Control target 4 × 105 ion counts
Maximum ion time 50 ms
Scan range 350–1,800 m/z
dd-MS2
Microscans 1
Detector type Orbitrap
Orbitrap resolution 50,000 (TMT-10)
AGC target 1 × 105 ion counts
Maximum ion time 105 ms
Cycle time: time between master scans 2 s
Isolation window 0.7 m/z
Fixed first mass 110 m/z
Activation type Higher-energy collision dissociation (HCD)
HCD energy (%) dd settings 38 (TMT)
Monoisotopic peak determination Peptide
Charge exclusion Include 2–6 charge states
Exclude isotopes On
Dynamic exclusion 20 s