Table 4. Parameters, reference values and comments of ultrasound assessment of breathing effort.
Technique | Parameter | Reference values | Comments |
---|---|---|---|
Diaphragm thickness (Tdi) | Diaphragm thickness at end-expiration | 1.5–2.4 mm (55,56) at end-expiration | Non-invasive and useful to assess development of atrophy. Not a measurement of effort per se. Left hemidiaphragm may be difficult to visualize |
Thickening fraction (TFdi) | TFdi = (Tdi,ei − Tdi,ee)/Tdi,ee | 24–53% during quiet breathing (57), up to 157% during vigorous effort (58) | Does not allow for direct quantification of muscle pressures |
Caudal displacement | M-mode of diaphragm displacement during tidal breathing | 1.6–1.8 cm during quiet breathing, up to 7.5 cm during deep breathing (59) | Cannot distinguish patient work from ventilator work during partially supported ventilation |
US, ultrasound; Tdi (ee/ei), diaphragm thickness (at end-expiration/end-inspiration); TFdi, thickening fraction of the diaphragm.