Table 2.

Assessment of precision and accuracy of α-, β-, and γ-amanitin with and without internal standards (MetSSP and 15N10 α-amanitin) for QC Low (7.0 ng/mL) and QC Medium (75 ng/mL) in pooled urine for six calibration curves.

α-amanitin
Internal Standard
None MetSSP 15N10-α-amanitin
QC Low QC Medium QC Low QC Medium QC Low QC Medium
Mean 7.20 79.4 7.53 85.3 6.95 78.4
CV 13.0% 8.75% 17.5% 8.01% 3.96% 1.51%
Accuracy 103% 107% 108% 114% 99.3% 105%
β-amanitin
Internal Standard
None MetSSP 15N10-α-amanitin
QC Low QC Medium QC Low QC Medium QC Low QC Medium
Mean 6.99 77.3 7.07 84.5 6.76 76.4
CV 12.4% 8.45% 10.4% 15.9% 16.4% 16.7%
Accuracy 99.9% 103% 101% 113% 96.6% 102%
ɣ-amanitin
Internal Standard
None MetSSP 15N10-α-amanitin
QC Low QC Medium QC Low QC Medium QC Low QC Medium
Mean 7.05 79.2 7.47 85.6 6.97 77.6
CV 12.7% 10.9% 16.3% 14.9% 13.0% 14.3%
Accuracy 101% 106% 106% 115% 99.6% 104%