Skip to main content
. Author manuscript; available in PMC: 2019 Nov 1.
Published in final edited form as: Comput Biol Med. 2018 May 8;102:242–250. doi: 10.1016/j.compbiomed.2018.05.002

Table 4.

Logistic regression model for prediction of S-ICD screening failure

Supine 12-lead ECG parameter Beta-coefficient
R-wave amplitude in lead II, μV 2.418
S-wave amplitude in lead II, μV −0.747
S-wave amplitude in aVL, μV −2.485
T-wave amplitude in lead I, μV 58.411
T-wave amplitude in lead II, μV −94.181
T-wave amplitude in lead aVL, μV −18.308
T-wave amplitude in lead aVF, μV 89.520
T-wave amplitude in V3, μV −3.476
T-wave amplitude in V4, μV 5.280
T-wave amplitude in V5, μV −2.800
T-wave amplitude in V6, μV −3.097
R/Tmax −60.729
Constant(intercept) 300.692